Laboratory Evaluation of Hemostasis



Laboratory Evaluation of Hemostasis


George A. Fritsma



Objectives


After completion of this chapter, the reader will be able to:


1. Properly collect and transport hemostasis blood specimens.


2. Reject hemostasis blood specimens due to clots, short draws, or hemolysis.


3. Prepare hemostasis blood specimens for analysis.


4. Describe the principles of platelet aggregometry.


5. Apply appropriate platelet function tests in a variety of conditions and interpret their results.


6. Diagnose von Willebrand disease and monitor its treatment.


7. Analyze plasma markers of platelet activation platelet factor 4 and β-thromboglobulin.


8. Describe the principle of, appropriately select, and correctly interpret the results of clot-based coagulation screening tests, including activated clotting time, prothrombin time, partial thromboplastin time, and the thrombin clotting time.


9. Interpret clot-based screening test results collectively to reach presumptive diagnoses, then recommend and perform confirmatory tests.


10. Perform partial thromboplastin time mixing studies to detect factor deficiencies, lupus anticoagulants, and specific factor inhibitors.


11. Describe the principle of, appropriately select, and correctly interpret coagulation factor assays.


12. Describe the principle of and correctly interpret Bethesda titers for coagulation factor inhibitors.


13. Describe the principle of, appropriately select, and correctly interpret tests of fibrinolysis, including assays for D-dimer, plasminogen, plasminogen activators, and plasminogen activator inhibitors.



Case Study


After studying the material in this chapter, the reader should be able to respond to the following case study:


A 54-year-old woman experienced a pulmonary embolism on September 26 and began oral anticoagulant therapy. Monthly PT values were collected to monitor therapy. From October through January, her INR was stable at 2.4, but on February 1 her INR was 1.3. The reduced INR was reported to her physician.


On questioning, the patient reported that there had been no change in her warfarin (Coumadin) dosage or in her diet. She recalled, however, that the phlebotomist had used a tube with a red and black stopper. She had thought this to be out of the ordinary and had remarked about it to the phlebotomist, who made no response. The medical laboratory scientist who had performed the PT assay reexamined the blood specimen and saw that it was in a blue-topped tube.




Hemostasis Specimen Collection


Most hemostasis laboratory procedures are performed on venous whole blood collected by venipuncture and mixed 9 : 1 with a 3.2% solution of sodium citrate anticoagulant. The specimen is maintained as well-mixed whole blood for platelet function testing or centrifuged to provide platelet-poor plasma (PPP) for other procedures. Phlebotomists, patient care technicians, nurses, medical laboratory scientists, and other health personnel who collect blood specimens must adhere closely to published protocols for specimen collection and management. The nursing or laboratory supervisor is responsible for the current validity of specimen collection and handling protocols and ensures that personnel employ approved techniques.1



Patient Management During Hemostasis Specimen Collection


Patients need not fast for hemostasis testing, and little special preparation is required before the collection of hemostasis laboratory specimens. However, there are numerous drugs that affect the outcomes of coagulation tests; for example, aspirin suppresses most platelet function and warfarin (Coumadin) reduces the activities of factor II (prothrombin), factor VII, factor IX, factor X, protein C, protein S, and protein Z and prolongs the prothrombin time (PT). The phlebotomist should attempt to record all drugs the patient is currently taking, and patients should be instructed by their physicians to discontinue drugs that may interfere with coagulation test results before testing.


Phlebotomists may manage patients using standard protocols for identification, cleansing, tourniquet use, and venipuncture (see Chapter 3). If there is a reason to anticipate excessive bleeding—for instance, if the patient has multiple bruises or mentions a tendency to bleed—the phlebotomist should extend the time for observing the venipuncture site from 1 minute to 5 minutes and should apply a pressure bandage before dismissing the patient.



Hemostasis Specimen Collection Tubes


Most hemostasis specimens are collected in plastic blue-stopper (blue-top, blue-closure) sterile evacuated blood collection tubes containing a measured volume of 0.105 to 0.109 mol/L (3.2%) buffered sodium citrate anticoagulant.2 Tubes of uncoated soda-lime glass are unsuitable because their negative surface charge activates platelets and plasma procoagulants. Siliconized (plastic-coated) glass tubes are available, but their use is waning because of concern for potential breakage with consequent risk of exposure to bloodborne pathogens.3



Hemostasis Specimen Collection Rules




• If the hemostasis specimen is part of a series of tubes to be filled from a single venipuncture site, it must be collected first or immediately after a nonadditive tube. The hemostasis tube may not immediately follow a tube that contains heparin (green stopper), ethylenediamine tetraacetic acid (EDTA, lavender stopper), sodium fluoride (gray stopper), or clot-promoting silica particles such as are contained in plastic red-topped or serum separator (gel) tubes. These additives may become transferred to the hemostasis specimen on the stopper needle and invalidate all hemostasis test results. Nonadditive tubes includes red-topped glass tubes and clear-topped or red and gray marble–topped tubes. If nonadditive tubes are unavailable, the phlebotomist may use and discard a preliminary blue-topped tube.4


• The ratio of whole blood to anticoagulant must be 9 parts blood to 1 part anticoagulant. Evacuated tubes are designed so that the negative internal pressure draws the correct volume of blood from the vein. Collection tube manufacturers indicate the allowable range of collection volume error in package inserts and provide a minimum volume line on each tube. In most cases, the volume of blood collected must be within 90% of the calibrated volume. A short draw—that is, a specimen with a smaller volume than that specified by the manufacturer—generates erroneously prolonged clot-based coagulation test results because the excess anticoagulant relative to blood volume neutralizes test reagent calcium.5 Short-draw specimens are consistently discarded, and a fresh specimen is collected from the patient. Most plastic blue-topped tubes collect 3.0 mL of whole blood; the smaller the collection tube, the narrower the tolerance for short draws.


• When specimens are collected using winged-needle butterfly sets, the phlebotomist must compensate for the internal volume of the tubing, which is usually 12 inches long and contains approximately 0.5 mL of air. The phlebotomist must first collect and discard a nonadditive tube or an identical blue-topped tube. This step ensures that the needle set tubing is filled with fresh patient blood before the hemostasis specimen is collected.6


• Clotted specimens are useless for hemostasis testing, even if the clot is small. A few seconds after collection the phlebotomist must gently invert the specimen at least five times to mix the blood with the anticoagulant and prevent clot formation. If possible, the medical laboratory scientist must visually examine for clots just before centrifugation and testing. Many coagulometers are equipped to detect the presence of clots. Clotted specimens are discarded, and a new specimen is collected from the patient.


• Excessive specimen agitation causes hemolysis, procoagulant activation, and platelet activation. The phlebotomist must never shake the tube. The test results from visibly hemolyzed specimens are unreliable, and the specimen must be recollected (Table 45-1).7



• Excess manipulation of the needle may promote the release of procoagulant substances from the skin and connective tissue, which contaminate the specimen and cause clotting factor activation. Consequently, test results from specimens collected during a traumatic venipuncture may be falsely shortened and unreliable.8


• During blood collection, the phlebotomist must remove the tourniquet within 1 minute of its application to avoid stasis.9 Stasis is a condition in which venous flow is slowed. Stasis results in the local accumulation of coagulation factor VIII and von Willebrand factor (VWF), which may result in false shortening of clot-based coagulation test results.


Managers of many hemostasis specialty laboratories insist that specimens from patients from whom it is difficult to draw blood and specimens for specialized tests such as platelet aggregometry be collected by syringe, as described in the following sections. Many hemostasis laboratories employ medical laboratory scientists and phlebotomists who are specially trained in specimen collection to ensure the integrity of the specimen.



Specimen Collection Using Syringes and Winged-Needle Sets


Although syringes are impractical for high-volume hemostasis screening, they are occasionally used in place of evacuated blood collection units. Syringes, joined with winged-needle sets, are especially useful for collecting specimens from patients whose veins are small, fragile, or scarred by repeated venipunctures. The use of syringes presents additional needle stick risk to the phlebotomist, so careful training and handling are essential.10


The phlebotomist selects sterile syringes of 20-mL capacity or less with nonthreaded Luer-slip hubs. The phlebotomist assembles syringes, a winged needle set (Figure 45-1), a tubing clamp, and standard venipuncture materials. The phlebotomist then uses the following protocol:




1. Use standard patient identification and standard blood specimen management precautions (see Chapter 3)


2. Most syringes are delivered with the plunger withdrawn about 1 mm from the end of the barrel. Move the plunger outward and inward within the barrel. Expel all air from the barrel and affix the needle set to the Luer-slip hub.


3. Optional: draw precisely measured anticoagulant into the syringe prior to collection.


4. Cleanse the venipuncture site, affix the tourniquet, and insert the winged needle. Immobilize the needle set by loosely taping the tube to the arm about 2 inches from the needle.


5. Fill the syringe using a gentle, even pressure.


6. Place the syringe on a clean surface and clamp the tubing with a hemostat near the needle hub.


7. Attach a second syringe if needed; release the clamp and fill the second syringe.


8. Replace the clamp, remove the needle set, and immediately activate the needle cover.


After seeing to the patient’s welfare, the phlebotomist cautiously transfers the blood specimen to sealed evacuated tubes by affixing a safety transfer device. The specimen is allowed to flow gently down the side of the tube. The specimen is not pushed forcibly into the tube, because agitation causes hemolysis and platelet activation. The transfer must be accomplished within a few seconds of the time the syringe is filled, and the tube must be gently inverted at least five times. The specimen volume must be correct.



Selection of Needles for Hemostasis Specimens


Whether evacuated collection tubes or syringes are used, the bore of the needle should be sufficient to prevent hemolysis and activation of platelets and plasma procoagulants. If the overall specimen is 25 mL, a 20- or 21-gauge thin-walled needle is used (Table 45-2). For a larger specimen, a 19-gauge needle is required. A 23-gauge needle is acceptable for pediatric patients or patients whose veins are small, but the negative collection pressure must be reduced. All needles provide safety closures that either cover or blunt the needle immediately after completion of the venipuncture.




Collection of Specimens from Vascular Access Devices


Blood specimens may be drawn from heparin or saline locks, ports in intravenous lines, peripherally inserted central catheters (PICC tubes), central venous catheters, or dialysis catheters. Vascular access device management requires strict adherence to protocol to ensure sterility, prevent emboli, and prevent damage to the device. Personnel must be trained and must recognize the signs of complications and take appropriate action. Institutional protocol may limit vascular access device blood collection to physicians and nurses. Before blood is collected for hemostasis testing, the line must be flushed with 5 mL of saline, and the first 5 mL of blood, or six times the volume of the tube, must be collected and discarded. The phlebotomist must not flush with heparin. Blood is collected into a syringe and transferred to an evacuated tube as described in the prior section on hemostasis specimen collection with syringes and winged needle sets.11



Specimen Collection Using Capillary Puncture


Several near-patient testing (point-of-care) coagulometers (see Chapter 47) generate PT results from a specimen consisting of 10 to 50 mcL of whole blood. These instruments are designed to test either anticoagulated venous whole blood or capillary (finger-stick) blood and represent a significant convenience to patients and to anticoagulation clinics.12 Many are designed for patient self-testing, and pediatric or neonatal testing, and laboratory scientists are often charged with training patients in proper capillary puncture technique.13


Capillary specimen punctures are made using sterile spring-loaded lancets designed to make a cut of standard depth and width while avoiding injury (see Chapter 3). The phlebotomist or patient selects and cleanses the middle or fourth (ring) finger and activates the device so that it produces a puncture that is just off center of the fingertip and perpendicular to the fingerprint lines. After wiping away the first drop of blood, which is likely to be contaminated by tissue fluid, the phlebotomist places the collection device directly adjacent to the free-flowing blood and allows the device to fill. The phlebotomist wipes excess blood from the outside of the device and introduces it to the coagulometer to complete the assay. The phlebotomist then presses a gauze pad to the wound and instructs the patient to maintain pressure until bleeding ceases.


The key to accurate measurement of PT is a free-flowing puncture. Often it is necessary for the phlebotomist to warm the patient’s hand to increase blood flow to the fingertips. Blood collection device distributors provide dry disposable warming devices for this purpose. The phlebotomist avoids squeezing (“milking”) the finger, because this renders the blood specimen inaccurate by raising the concentration of tissue fluid relative to blood cells.14



Anticoagulants Used for Hemostasis Specimens


Sodium Citrate (Primary Hemostasis Anticoagulant)


The anticoagulant used for hemostasis testing is buffered 3.2% (0.105 to 0.109 mol/L) sodium citrate, Na3C6H5O7•2H2O, molecular weight 294.10. Sodium citrate binds calcium ions to prevent coagulation, and the buffer stabilizes specimen pH as long as the tube stopper remains in place.15


The anticoagulant solution is mixed with blood to produce a 9 : 1 ratio; 9 parts whole blood to 1 part anticoagulant. In most cases, 0.3 mL of anticoagulant is mixed with 2.7 mL of whole blood, which are the volumes in the most commonly used evacuated plastic collection tubes, but any volumes are valid, provided that the 9 : 1 ratio is maintained. The ratio yields a final citrate concentration of 10.5 to 10.9 mmol/L of anticoagulant in whole blood.16 Some laboratory scientists prepare specimen tubes locally for special hemostasis testing.



Adjustment of Sodium Citrate Volume for Elevated Hematocrits

The 9 : 1 blood-to-anticoagulant ratio is effective provided the patient’s hematocrit is 55% or less. In polycythemia, the decrease in plasma volume relative to whole blood unacceptably raises the anticoagulant-to-plasma ratio, which causes falsely prolonged results on clot-based coagulation tests. The phlebotomist must provide tubes with relatively reduced anticoagulant volumes for collection of blood from a patient whose hematocrit is known to be 55% or higher. The amount of anticoagulant needed may be computed for a 5-mL total specimen volume by using the graph in Figure 45-2 or the following formula, which is valid for any total volume:



C=(1.85×103)(100H)V


image

where C is the volume of sodium citrate in milliliters, V is volume of whole blood–sodium citrate solution in milliliters, and H is the hematocrit in percent.


For example, to collect 3 mL of blood and anticoagulant mixture from a patient who has a hematocrit of 65%, calculate the volume of sodium citrate as follows:


C=(1.85×103)(10065%)×3.0 mL


image

C=(1.85×103)(35%)×3.0 mL


image

C=0.19 mL of 3.2% sodium citrate


image

Remove the tube stopper and pipette and discard 0.11 mL from the 0.3 mL of anticoagulant, leaving 0.19 mL. Collect blood in a syringe and transfer approximately 2.89 (2.9) mL of blood to the tube; replace the stopper, and immediately mix by inverting four times. Some highly skilled laboratory scientists actually puncture the stopper with a tuberculin syringe equipped with a 25-gauge needle, invert the tube, and withdraw the computed volume, leaving the negative pressure intact. Alternatively, the laboratory scientist can prepare for collection of 10 mL of blood and anticoagulant solution in a 12-mL centrifuge tube as follows:


C=(1.85×103)(35%)×10.0 mL


image

C=0.64 mL of 3.2% sodium citrate


image

In this instance, 0.64 mL of sodium citrate is pipetted into the tube and 9.36 (9.4) mL of whole blood is transferred from the collection syringe.


There is no evidence suggesting a need for increasing the volume of anticoagulant for specimens from patients with anemia, even when the hematocrit is less than 20%.



Other Anticoagulants Used for Hemostasis Specimens


Few data support the use of EDTA-anticoagulated specimens for coagulation testing, because calcium ion chelation interferes with coagulation assays.17 EDTA is the anticoagulant used in collecting specimens for complete blood counts, including platelet counts. EDTA may be required for specimens used for molecular diagnostic testing, such as testing for factor V Leiden mutation or the prothrombin G20210A mutation. Likewise, acid citrate dextrose (ACD, yellow stopper) and dipotassium EDTA (K2EDTA) with gel (white stopper) tubes may be used for molecular diagnosis, as specified by institutional protocol. Heparinized specimens have never been validated for use in plasma coagulation testing but may be necessary in cases of platelet satellitosis (satellitism) as a substitute for specimens collected in EDTA or sodium citrate. Citrate theophylline adenosine dipyridamole (CTAD, blue stopper) tubes are used to halt in vitro platelet or coagulation activation for specialty assays such as those for the platelet activation markers platelet factor 4 (PF4) and platelet surface membrane P-selectin (measured by flow cytometry) or the coagulation activation markers prothrombin fragment 1+2 and thrombin-antithrombin.



Hemostasis Specimen Management


Hemostasis Specimen Storage Temperature


Sodium citrate–anticoagulated whole-blood specimens are placed in a rack and allowed to stand in a vertical position with the stopper intact and uppermost. The pH remains constant as long as the specimen is sealed. Specimens are maintained at 18° C to 24° C (room temperature), never at refrigerator temperatures (Table 45-3). Storage at 1° C to 6° C activates factor VII, activates platelets, and causes the cryoprecipitation of large VWF multimers.18,19 Also, specimens should never be stored at temperatures greater than 24° C, because heat causes deterioration of coagulation factor VIII.





Preparation of Hemostasis Specimens for Assay


Whole-Blood Specimens Used for Platelet Aggregometry


Blood for whole-blood platelet aggregometry or lumiaggregometry must be collected with 3.2% sodium citrate and held at 18° C to 24° C until testing. Chilling destroys platelet activity. Aggregometry may be started immediately and must be completed within 3 hours of specimen collection. The scientist mixes the specimen by gentle inversion, checks for clots just before testing, and rejects specimens with clots. Most specimens for whole-blood aggregometry are mixed 1 : 1 with normal saline before testing, although if the platelet count is less than 100,000/mcL the specimen is tested undiluted.21



Platelet-Rich Plasma Specimens Used for Platelet Aggregometry


Light-transmittance (optical) platelet aggregometers are designed to test platelet-rich plasma (PRP), plasma with a platelet count of 200,000 to 300,000/mcL. Sodium citrate–anticoagulated blood is first checked visually for clots and then centrifuged at 50 g for 30 minutes with the stopper in place to maintain the pH. The supernatant PRP is transferred by a plastic pipette to a clean plastic tube, and the tube is sealed and stored at 18° C to 24° C (room temperature) until the test is begun. PRP-based light-transmittance aggregometry is initiated no less than 30 minutes after the specimen is centrifuged and completed within 3 hours of the time of collection. To produce sufficient PRP, the original specimen must measure 9 to 12 mL of whole blood. Light-transmittance aggregometry is unreliable when the patient’s whole-blood platelet count is less than 100,000/mcL.



Platelet-Poor Plasma Required for Clot-Based Testing


Clot-based plasma coagulation tests require PPP–plasma with a platelet count of less than 10,000/mcL.22 Sodium citrate–anticoagulated whole blood is centrifuged at 1500 xg for 15 minutes in a swinging bucket centrifuge to produce supernatant PPP. Alternatively, the angle-head StatSpin Express 2 (Iris Sample Processing, Inc., Westwood, Mass.) generates 4400 xg and can produce PPP within 3 minutes. Both make it possible for automated coagulometers to sample from the supernatant plasma of the primary tube. The advantage of the slower swinging bucket centrifuge head is that it produces a straight, level plasma–blood cell interface, whereas angle-head centrifuge heads cause platelets to adhere to the side of the tube. If the tube is allowed to stand, the platelets drift back into the plasma and release granule contents. Each hemostasis laboratory manager establishes the correct centrifugation speed and times for the local laboratory, and centrifugation must yield PPP from specimens with high initial platelet counts.


In the special hemostasis laboratory the manager may choose a double-spin approach. The primary tube is centrifuged using a swinging bucket centrifuge, and the plasma is transferred to a secondary tube, which is labeled and centrifuged again. The double-spin approach may be used to produce PPP with a plasma platelet count of less than 5000/mcL, which is preferred for lupus anticoagulant (LA) testing and for preparation of frozen plasma.


The presence of greater than 10,000 platelets/mcL in plasma affects clot-based test results. Platelets are likely to become activated in vitro and release the membrane phospholipid phosphatidylserine, which triggers plasma coagulation and neutralizes LA if present, interfering with LA testing. Platelets also secrete fibrinogen, factors V and VIII, and VWF (see Chapter 13). These may desensitize PT and PTT assays and interfere with clot-based coagulation assays. In addition, platelets release PF4, a protein that binds and neutralizes therapeutic heparin in vitro, falsely shortening the PTT and interfering with heparin management.


The hemostasis laboratory manager arranges to perform plasma platelet counts on coagulation plasmas at regular intervals to ensure that they are consistently platelet poor. Many managers select 10 to 12 specimens from each centrifuge every 6 months, perform platelet counts, and document that their samples remain appropriately platelet poor, even if the initial platelet count is elevated.


Laboratory scientists inspect hemostasis plasmas for hemolysis (red), lipemia (cloudy, milky), and icterus (golden yellow from bilirubin). Visible hemolysis implies platelet or coagulation pathway activation. Visibly hemolyzed specimens are rejected, and new specimens must be obtained. Lipemia and icterus may affect the end-point results of optical coagulation instruments. The hemostasis laboratory manager may choose to maintain a separate mechanical end-point coagulometer to substitute for the optical instrument if the specimen is too cloudy for optical determinations. Conversely, some optical instruments detect and compensate for lipemia and icterus via spectrophotometric analysis.23



Specimen Storage


Specimens for PT assay only may be held uncentrifuged at 18° C to 24° C for up to 24 hours provided the tubes remain closed. Likewise, specimens for PTT measurement may be held uncentrifuged for up to 4 hours. However, specimens from patients receiving unfractionated heparin collected for PTT heparin monitoring must be centrifuged and the supernatant PPP sampled or transferred within 1 hour to avoid false shortening of the PTT as PF4 neutralizes the heparin.


If the hemostasis test cannot be completed within the prescribed interval, the laboratory scientist must immediately centrifuge the specimen. The supernatant PPP must be transferred by plastic pipette to a plastic freezer tube, sealed, and frozen, and may be stored at −20° C for up to 2 weeks or at −70° C for up to 6 months. At the time the test is performed, the specimen must be thawed rapidly at 37° C, mixed well, and tested within 1 hour of the time it is removed from the freezer. If it cannot be tested immediately, the specimen may be stored at 1° C to 6° C for 2 hours after thawing. To avoid cryoprecipitation of VWF, specimens may not be frozen and thawed more than once.



Platelet Function Tests


Platelet function tests are designed to detect qualitative (functional) platelet abnormalities in patients who are experiencing the symptoms of mucocutaneous bleeding (see Chapter 44). A platelet count is performed and the blood film is reviewed before platelet function tests are begun, because thrombocytopenia is a common cause of hemorrhage (see Chapter 43).24 Qualitative platelet abnormalities are suspected only when bleeding symptoms are present and the platelet count exceeds 50,000/mcL.


Although hereditary platelet function disorders are rare, acquired defects are common.25 Acquired platelet defects are associated with liver disease, renal disease, myeloproliferative neoplasms, myelodysplastic syndromes, myeloma, uremia, autoimmune disorders, anemias, and drug therapy. Platelet morphology is often a clue; for instance, in Bernard-Soulier syndrome the blood film reveals mild thrombocytopenia and large gray platelets (see Figure 44-1). Similarly, the presence of large platelets on the blood film associated with elevated mean platelet volume often indicates rapid platelet turnover, such as occurs in immune thrombocytopenic purpura or thrombotic thrombocytopenic purpura. Giant or dysplastic platelets are seen in myeloproliferative neoplasms, acute leukemia, and myelodysplastic syndromes.



Bleeding Time Test for Platelet Function


The bleeding time test was the original test of platelet function, although it is now largely replaced by near-patient analysis of platelet function using the PFA-100 (Siemens Healthcare Diagnostics, Inc., Deerfield, Ill.), the Ultegra (Accumetrics, San Diego, Calif.), or platelet aggregometry.26 To perform the test, the phlebotomist uses a lancet to make a small controlled puncture wound and records the duration of bleeding, comparing the results with the universally accepted reference interval of 2 to 9 minutes. The bleeding time test was first described by Duke27 in 1912 and modified by Ivy28 in 1941. In 1978 some standardization was attempted: a blood pressure cuff was inflated to 40 mm Hg, a calibrated spring-loaded lancet (Surgicutt Bleeding Time Device; International Technidyne Corporation, Edison, N.J.) was triggered on the volar surface of the forearm a few inches distal to the antecubital crease, and the resulting wound was blotted every 30 seconds with filter paper until bleeding stopped.29,30


A prolonged bleeding time could theoretically signal a functional platelet disorder such as von Willebrand disease (VWD) or a vascular disorder such as scurvy or vasculitis, and was a characteristic result of therapy with aspirin and other nonsteroidal antiinflammatory drugs (NSAIDs). Measurement of the bleeding time was often requested by surgeons at admission in an attempt to predict surgical bleeding, but a series of studies in the 1990s revealed that the test has inadequate predictive value. The bleeding time is affected by the nonplatelet variables of intracapillary pressure, skin thickness at the puncture site, and size and depth of the wound, all of which interfere with accurate interpretation of the test results. Owing to its poor predictive value for bleeding and its tendency to scar the forearm, use of the bleeding time assay has been discontinued at most institutions.



Platelet Aggregometry and Lumiaggregometry


Functional platelets adhere to subendothelial collagen, aggregate with each other, and secrete the contents of their α-granules and δ-granules (see Chapter 13). Normal adhesion requires intact platelet membranes and functional plasma VWF. Normal aggregation requires that platelet membranes and platelet activation pathways are intact, that the plasma fibrinogen concentration is normal, and that normal secretions are release from platelet granules. Platelet adhesion, aggregation, and secretion are assessed using in vitro platelet aggregometry.


An aggregometer is an instrument designed to measure platelet aggregation in a suspension of citrated whole blood or PRP. Specimens are collected and managed in compliance with standard laboratory protocol as described in the section on preparation of hemostasis specimens for assays, and maintained at room temperature (18° C to 24° C) until testing begins. Specimens for PRP-based light-transmittance aggregometry must stand undisturbed for 30 minutes after centrifugation while the platelets regain their responsiveness. Specimens for impedance aggregometry are diluted 1 : 1 with normal saline and tested immediately. Specimens must be tested within 3 hours of collection to avoid spontaneous in vitro platelet activation. Platelet aggregometry is a high complexity laboratory test requiring a skilled, experienced operator.



Platelet Aggregometry Using Platelet-Rich Plasma


PRP aggregometry is performed using a specialized photometer called a light-transmittance aggregometer (PAP-8E Platelet Aggregation Profiler; Bio/Data Corporation, Horsham, Pa.).31 After calibrating the instrument in accordance with manufacturer instructions, the operator pipettes the PRP to instrument-compatible cuvettes, usually 500 mcL; drops in one clean plasticized stir bar per sample; places the cuvettes in incubation wells; and allows the samples to warm to 37° C for 5 minutes. The operator then transfers the first cuvette, containing specimen and stir bar, to the instrument’s reaction well and starts the stirring device and the recording computer. The stirring device turns the stir bar at 800 to 1200 rpm, a gentle speed that keeps the platelets in suspension. The instrument directs focused light through the sample cuvette to a photodetector (Figure 45-3). As the PRP is stirred, the recorder tracing first stabilizes to generate a baseline, near 0% light transmission. After a few seconds, the operator forcibly pipettes an agonist (aggregating agent) into the sample to trigger aggregation. In a normal specimen, after the agonist is added, the shape of the suspended platelets changes from discoid to spherical, and the intensity of light transmission increases in proportion to the degree of shape change. Percent transmittance is monitored continuously and recorded (Figure 45-4). As platelet aggregates form, more light passes through the PRP, and the tracing begins to move toward 100% light transmission. Platelet deficiencies are reflected in diminished or absent aggregation; many laboratory directors choose 40% aggregation as the lower limit of normal.





Whole-Blood Platelet Aggregometry


In whole-blood platelet aggregometry, platelet aggregation is measured by electrical impedance using a 1 : 1 saline–whole blood suspension (Model 700 Whole Blood/Optical Lumi-Aggregometer; Chrono-log Corporation, Havertown, Pa.).32 The operator pipettes aliquots of properly mixed whole blood to cuvettes and adds equal volumes of physiologic saline. Suspension volume may be 300 to 500 mcL. The operator drops in one stir bar per cuvette and places the cuvettes in 37° C incubation wells for 5 minutes. The operator transfers the first cuvette to a reaction well, forcibly pipettes in an agonist, and suspends a pair of low-voltage cartridge-mounted disposable direct current (DC) electrodes in the mixture. As aggregation occurs, platelets collect on the electrodes, impeding the DC current (Figure 45-5). The rise in impedance is directly proportional to platelet aggregation and is amplified and recorded by instrument circuitry. A whole-blood aggregometry tracing closely resembles a PRP-based light-transmittance aggregometry tracing as shown in Figure 45-4.




Platelet Lumiaggregometry


The Whole Blood/Optical Lumi-Aggregometer may be used for simultaneous measurement of platelet aggregation and the secretion of adenosine triphosphate (ATP) from activated platelet δ-granules.33 The procedure for lumiaggregometry differs little from that for conventional aggregometry and simplifies the diagnosis of platelet dysfunction.34 As ATP is released it oxidizes a firefly-derived luciferin-luciferase reagent (Chrono-lume; Chrono-log Corporation) to generate cold chemiluminescence proportional to the ATP concentration. A photodetector amplifies the luminescence, which is recorded as a second tracing on the aggregation report.35


Lumiaggregometry may be performed using whole blood or PRP.36 To perform lumiaggregometry, the operator adds an ATP standard to the first sample, then adds luciferin-luciferase and tests for full luminescence. The operator then adds luciferin-luciferase and an agonist to the second sample; the instrument monitors for aggregation and secretion simultaneously. Thrombin is typically the first agonist used, because thrombin induces full secretion. The luminescence induced by thrombin is measured, recorded, and used for comparison with the luminescence produced by the additional agonists. Normal secretion induced by agonists other than thrombin produces luminescence at a level of about 50% of that resulting from thrombin. Figure 45-6 depicts simultaneous aggregation and secretion responses to thrombin; Figure 45-7 is a scanning electron micrograph of resting and activated platelets.





Platelet Agonists (Activating Agents) Used in Aggregometry


The agonists used most frequently in clinical practice are thrombin or synthetic thrombin receptor–activating peptide (TRAP), adenosine diphosphate (ADP), epinephrine, collagen, arachidonic acid, and ristocetin. Table 45-4 lists representative concentrations and platelet activation pathways tested by each agonist. Small volumes (2 to 5 mcL) of concentrated agonist are used so that they have little dilutional effect in the reaction system.37



Thrombin (or TRAP) cleaves two platelet membrane protease-activatable receptors (PARs), PAR1 and PAR2, both members of the seven-transmembrane repeat receptor family (see Chapter 13). Thrombin or TRAP also cleaves glycoprotein (GP) 1bα and GP V. Internal platelet activation is effected by membrane-associated G proteins and both the eicosanoid and the diacylglycerol pathways. Thrombin-induced activation results in full secretion and aggregation. In lumiaggregometry, the operator ordinarily begins with 1 unit/mL of thrombin or TRAP to induce the release of 1 to 2 nmol/L of ATP, detected by the firefly luciferin-luciferase luminescence assay. Other agonists—for instance, 5 mcg/mL of collagen—induce the release of at most 0.5 to 1.0 nmol/L of ATP. Thrombin-induced secretion may be diminished to less than 1 nmol/L in storage pool deficiencies (see Chapter 44), but is relatively unaffected by membrane disorders or pathway enzyme deficiencies.


Reagent thrombin is stored dry at −20° C and is reconstituted with physiologic saline immediately before use. Leftover reconstituted thrombin may be divided into aliquots, frozen, and thawed for later use. Thrombin has the disadvantage that it often triggers coagulation simultaneously with aggregation. The use of TRAP avoids this pitfall.


ADP binds platelet membrane receptors P2Y1 and P2Y12, also members of the seven-transmembrane repeat receptor family. ADP-induced platelet activation relies on the physiologic response of membrane-associated G protein and the eicosanoid synthesis pathway. The end product of eicosanoid synthesis, thromboxane A2, raises cytosolic free calcium, which mediates platelet activation and induces secretion of ADP stored in δ-granules. The secreted ADP activates neighboring platelets.


ADP is the most commonly used agonist, particularly in aggregometry systems that measure only aggregation and not luminescence. The operator adjusts the ADP concentration to between 1 mcmol/L and 10 mcmol/L to induce “biphasic” aggregation (see Figure 45-4). At ADP concentrations near 1 mcmol/L, platelets achieve only primary aggregation, followed by disaggregation. The recording deflects from the baseline for 1 to 2 minutes and then returns to baseline. Primary aggregation involves shape change with formation of microaggregates, both reversible. Secondary aggregation is the formation of full platelet aggregates after release of platelet δ-granule ADP. At agonist ADP concentrations near 10 mcmol/L, there is simultaneous irreversible shape change, secretion, and formation of aggregates, resulting in a monophasic curve and full deflection of the tracing. ADP concentrations between 1 and 10 mcmol/L induce a biphasic curve: primary aggregation followed by a brief flattening of the curve called lag phase and then secondary aggregation.


Operators expend considerable effort to discover the ADP concentration that generates a biphasic curve with a visible lag phase, because the appropriate concentration varies among patients. This enables operators to use aggregometry alone to distinguish between membrane-associated platelet defects and storage pool or release defects.


Lumiaggregometry provides a clearer and more reproducible measure of platelet secretion, rendering the quest for the biphasic curve unnecessary. Secretion in response to ADP at 5 mcmol/L is diminished in platelet membrane disorders, eicosanoid synthesis pathway enzyme deficiencies, storage pool deficiency, or aspirin, NSAID, or clopidogrel therapy.


Reagent ADP is stored at −20° C, reconstituted with physiologic saline, and used immediately after reconstitution. Leftover reconstituted ADP may be aliquoted and frozen for later use.


Epinephrine binds platelet α-adrenergic receptors, identical to muscle receptors, and activates the platelet through the same metabolic pathways as reagent ADP. The results of epinephrine-induced aggregation match those of ADP except that epinephrine cannot induce aggregation in storage pool disorder or eicosanoid synthesis pathway defects no matter how high its concentration. Epinephrine does not work in whole-blood aggregometry.


Epinephrine is stored at 1° C to 6° C and reconstituted with distilled water immediately before it is used. Leftover reconstituted epinephrine may be aliquoted and frozen for later use.


Collagen binds GP Ia/IIa and GP VI, but induces no primary aggregation. After a lag of 30 to 60 seconds, aggregation begins, and a monophasic curve develops. Aggregation induced by collagen at 5 mcg/mL requires intact membrane receptors, functional membrane G protein, and normal eicosanoid pathway function. Loss of collagen-induced aggregation may indicate a membrane abnormality, storage pool disorder, release defect, or the presence of aspirin.


Most managers purchase lyophilized fibrillar collagen preparations such as Chrono-Par Collagen (Chrono-log Corporation). Collagen is stored at 1° C to 6° C and used without further dilution. Collagen may not be frozen.


Arachidonic acid

Stay updated, free articles. Join our Telegram channel

Jun 12, 2016 | Posted by in HEMATOLOGY | Comments Off on Laboratory Evaluation of Hemostasis

Full access? Get Clinical Tree

Get Clinical Tree app for offline access