Laboratory aspects of blood transfusion

22 Laboratory aspects of blood transfusion




Safe and effective blood transfusion requires the combined efforts of blood transfusion services, biomedical scientists and clinicians to ensure the highest standards are applied to all the systems in a complex process from ‘vein to vein’. This chapter provides a description of the laboratory framework required to provide the right blood components to the right patients at the right time. The increased awareness of what can go wrong with blood transfusion comes from a number of sources including the Serious Hazards of Transfusion UK haemovigilance scheme, which was started in 1996.1,2 This confidential reporting scheme, using detailed root cause analysis of errors, has provided data that have informed both national bodies and local transfusion services of measures to introduce in order to reduce risk (Fig. 22.1). It is clear that multiple errors can contribute to a single adverse event and that many of these are outside the control of the transfusion laboratory.



Within the laboratory setting, the application of strict protocols for sample labelling and testing, robust laboratory procedures, reliable documentation, frequent staff training and competency assessment should be used. Recently, the UK Transfusion Collaborative, comprising representatives of the professional bodies involved in UK transfusion practice, has produced some recommendations to reduce laboratory errors in transfusion. These address issues such training, competency, staffing levels and automated systems.3


This chapter is concerned with the testing of patient samples prior to the provision of appropriate compatible blood components including identification of red cell antibodies. It also covers compatibility testing and investigation of transfusion reactions and the testing required in other special situations including the antenatal and postnatal settings. National professional bodies such as the British Committee for Standards in Haematology (BCSH)4 and the AABB5 issue guidance to transfusion laboratories and this has been referenced where appropriate.


There is a new regulatory framework governing hospital transfusion laboratory practice which was implemented after the publication of two European Union Directives: 2002/98/EC and 2004/33/EC.6 In the UK these are the Blood Safety and Quality Regulations 2005 (Statutory Instruments 2005/50, 2005/1098 and 2006/2013) and they set standards for quality and safety of human blood and blood components in hospital ‘blood banks’ as well as ‘blood establishments’ (the UK Blood Services).7 ‘Blood banks’ are regulated in their roles of storing, distributing and performing compatibility tests, on blood and blood components for use in hospitals. The implications for hospital transfusion practice are the requirement for ‘vein to vein’ traceability of blood components, the importance of maintaining the ‘cold chain’ for all therapeutic blood components, the need to store transfusion records for 30 years and the requirement for a quality management system.6,7 UK laboratories have to assess themselves and submit an annual compliance report to the competent authority, which is the Medicines and Healthcare Products Regulatory Agency (MHRA).8 The MHRA carries out laboratory inspections to assess compliance with these regulations.


The UK government, through the National Blood Transfusion Committee, continues to promote the safe and effective transfusion of blood components and has published three Health Service Circulars.9 These documents aim to ensure that a team approach to blood transfusion safety is taken via the local clinical governance arrangements and to promote good transfusion practice via Hospital Transfusion Committees and Hospital Transfusion Teams supported by Regional Transfusion Committees.



Technology and automation in blood transfusion laboratories


Important changes have taken place in the blood transfusion laboratory in the last 10–15 years, as outlined in the previous edition of this book. As a result transfusion laboratory practice is safer, largely as a result of the implementation of new technologies for testing, automated systems to replace manual systems and the widespread use of information technology (IT) systems to support transfusion laboratory practice.


Barcoded labels on blood components, reagents, patient samples and equipment are now commonplace and these result in safer transfer of information, free from the transcription errors associated with manual methods.


Column agglutination (CAT) and solid-phase technology can both be used on automated machines and CAT can also be used by manual techniques. In UK hospital transfusion laboratories these technologies have replaced tube techniques and liquid-phase microplates for antibody screening and crossmatching (Figs 22.2, 22.3).10




The currently available CAT systems include Ortho BioVue, which comprises a six-well cassette containing a glass microbead matrix, ID DiaMed cards utilizing Sephadex gel matrix, also with six wells, and an eight-well Sephadex gel matrix card from Grifols. Each offers a wide range of profiles and reagent systems.


The currently available solid-phase systems are Immucor Capture-R and Bio-Rad Solidscreen II. The Immucor system utilizes a range of red cell antigens adhered to the surface of a U-shaped microplate well. Sensitized red cells are then used as a marker. The Bio-Rad Solidscreen II is a solid-phase method for antibody screening and identification. The wells of the microplates are coated with protein A to allow the reaction to bind to the plate forming a solid-phase cell layer.


Individual laboratories need to make careful and informed decisions when selecting reagents for pre-transfusion testing. It is vital that any abbreviated testing in an automated or semiautomated system is carefully evaluated for the risks that could ensue if important controls were omitted when using this technology for blood grouping.11


Laboratory information management systems (LIMS) store patient details and results of laboratory tests, allowing timely and accurate access to important information. In the transfusion department, IT systems have a much broader use and the updated BCSH guidelines for the specification and use of Information Technology (IT) systems in blood transfusion practice (2006) reflect this.12 Where possible, using bidirectional or unidirectional interfaces to automated blood grouping analysers, IT systems are used to eliminate errors that can arise when a manual step is employed, including interpretation of test results.


Computer algorithms support the ‘electronic issue’ of blood to patients with a negative antibody screen without the need to perform an antiglobulin crossmatch and, in the UK in 2009, 46% of laboratories were using this system for some or all of their patients (Fig. 22.4). The use of automation for all aspects of compatibility testing is now recommended practice as it is recognized that it is safer.3 Automation brings several or all of the discrete activities of compatibility testing into a single platform process. It provides various levels of increased security over manual testing and may provide justification for abbreviated pre-transfusion testing (e.g. abandoning duplicate D, previously termed ‘RhD’, typing or reverse ABO grouping in the presence of a valid historical group). A risk assessment must be made and documented prior to any abbreviation of an established procedure, with consideration being given to the presence or absence of key functions in the automated equipment. The BCSH guidelines for compatibility procedures12 and guidance from the MHRA13 give a list of factors to be taken into consideration and the reader is advised to consult these prior to implementing automated or semiautomated systems.




Pre-transfusion compatibility systems


The process of providing blood for transfusion involves many steps, all of which have to be reliably completed. These include the following:


Blood samples have to be taken from the correct patient and labelled at the bedside in a single uninterrupted procedure. The sample must be identified by four core patient identifiers: the correctly spelled forename and surname, the exact date of birth and an accurate unique patient number (such as the NHS number or equivalent). The sample should be dated and signed by the person taking the sample.14 Some guidelines recommend that addressograph labels should not be accepted on blood samples for compatibility testing because it increases the risk of ‘wrong blood in tube’(WBIT).15 In an international study from the BEST collaborative, it was estimated that miscollected samples demonstrating WBIT occurred at a median rate of 0.5 per 1000, although there was great variation worldwide in the reported frequency of mislabelled samples, probably resulting from variation in policies for sample acceptance.16 Labels printed at the bedside using barcoded patient wristband and hand-held scanners are acceptable.14 The laboratory should have a policy for rejecting badly labelled samples, although in 2004 a survey of hospitals in England and North Wales showed considerable variation in the content and application of these policies.17






Traceable documentation should exist to ensure that the results of laboratory compatibility procedures are available at the patient’s bedside to allow a check before transfusing the blood component. This should include a blood bag compatibility label (Fig. 22.5) and may include a compatibility form. The patient must be identified with a wristband and the blood component should be prescribed on a drug or fluid administration chart.14 In some countries, an additional bedside check of the patient’s blood group is undertaken prior to commencing the transfusion.



Documentation of the Transfusion Process


All stages of the transfusion process must be clearly documented and these records must be kept. In addition to guidance from the UK Royal College of Pathologists on the retention of documents,18 the BSQR 2005 regulations stipulate that the records must be accessible for 30 years.7 This allows any blood component to be traced from the donor to the recipient should information come to light about any potential infective risks to the recipient.


Computer records are easier to search than paper records, but laboratory information systems are likely to become obsolete and be replaced several times within this mandatory 30-year period, so provision must be made to store historical data in an accessible format when procuring a replacement computer system.12 Patient-held records are useful for patients who are treated in more than one institution, particularly if they have red cell antibodies and require phenotyped blood or if they have special requirements because of their underlying disease or its treatment. Credit card-sized records with corresponding patient information leaflets are issued by some transfusion centres to patients with red cell antibodies and similar cards exist for patients who require irradiated cellular blood components.19



Identification and Storage of Blood Samples


Depending on laboratory practice, blood transfusion tests use a clotted (serum) sample or EDTA-anticoagulated blood (plasma). Most laboratories using automated systems will use a plasma sample.


The reason for automated systems requiring plasma is because incompletely clotted blood samples may contain small fibrin clots that trap red cells into aggregates that could resemble agglutinates which could be falsely interpreted as a positive reaction. Clotted samples should be taken into a plain tube but not a tube with a separating gel.


The presence of complement in serum can cause lysis. If laboratory staff are used to recognizing agglutination as an indicator of a positive reaction they may fail to interpret the lysed red cells as an equally valid positive reaction. Therefore, when using serum for blood grouping and compatibility testing, any red cells in the test system should be washed and resuspended in saline which contains ethylenediamine tetra-acetic acid (EDTA) (see later). In addition, false-negative reactions may occur in immediate spin crossmatching with potent ABO antibodies, where rapid complement fixation causes a prozone effect (bound C1 inhibits agglutination). EDTA saline is not necessary when using plasma.



On being received in the laboratory, the details on the request form must be checked against the blood sample. Each blood sample must be labelled with a unique sample number. Barcode labels offer the advantage of positive sample identification and reduce the number of transcription errors. Samples inadequately or inaccurately labelled should NOT be used for pre-transfusion testing.11


Great care must be taken to select and identify the sample prior to any testing. Transposition of samples in the laboratory can lead to an incorrect blood group being assigned to a patient, with serious consequences, including ABO incompatible transfusions. Where possible, the primary sample should be used for testing. If samples are separated, the plasma must be clearly and accurately identified and special precautions must be taken. If repeated testing on this sample is anticipated, storing separate small aliquots reduces the risk of sample deterioration, which occurs with repeated thawing/freezing of larger samples.


Whole blood samples should be tested as soon as possible because they will deteriorate over time. Problems associated with storage include lysis of the red cells, loss of complement in the serum and decrease in potency of antibodies. The BCSH guidelines11 indicate working limits as outlined in Table 22.1. If separated plasma or serum samples are stored for later serological crossmatch, care must be taken to ensure that the patient has not been transfused in the interim. It has been recommended that samples should be kept for a minimum of 7 days from group and screen, stored at 4°C. Samples should be retained post-transfusion for investigation of acute transfusion reactions and preferably stored for 7 days post-transfusion to enable investigation of delayed transfusion reaction.18




ABO and D grouping


ABO and D grouping must be performed by a validated technique with appropriate controls. Before use, all new batches of grouping reagents should be checked for reliability by the techniques used in the laboratory. Grouping reagents should be stored according to manufacturer’s instructions.



ABO Grouping


ABO grouping is the single most important serological test performed in compatibility testing; consequently, it is imperative that the sensitivity and security of the test system are not compromised. The fact that anti-A and anti-B are naturally occurring antibodies allow the patient’s plasma to be tested against known A and B cells in a ‘reverse’ group.


This is an excellent built-in check for the ‘forward’ or cell group and has always been considered to be an integral part of ABO grouping, allowing the reading and recording of test results to be split into two discrete tasks. However, with secure, fully automated systems, linked to secure laboratory information management systems that, in combination, have the ability to prevent procedural ABO grouping errors, some laboratories now omit the reverse group when testing samples for which a historical group is available.11 This should only be considered following a careful risk assessment and taking into account that the first sample taken may have been from the wrong patient. This may be in the order of 1:2000 samples.16 Any discrepancy between the forward and reverse groups should be investigated further and any repeat tests should be undertaken using cells taken from the original sample rather than from a prepared cell suspension.




D Grouping


D grouping is usually undertaken at the same time as ABO grouping for convenience and to minimize clerical errors that may arise through repeated handling of patients’ samples. In the absence of secure automation, testing should be undertaken in duplicate because there is no counterpart of the ‘reverse’ grouping of ABO testing.



Reagents for D Grouping


Monoclonal reagents do not have the problem of possible contamination with antibodies of unwanted specificities, as was the case with polyclonal reagents. Therefore, the duplicate testing may be undertaken using the same anti-D reagent, although this should be dispensed as though it were two separate reagents.


DVI is the partial D with the fewest epitopes; therefore of all the D variants, DVI individuals are those most likely to form anti-D and a case of severe haemolytic disease of the fetus and newborn (HDFN) has been described.20 For this reason, anti-D monoclonal reagents that do not detect DVI should be selected for testing patients’ samples.21,22 The use of anti-CDE reagents has led to the misinterpretation of r′ and r″ cells in UK National External Quality Assessment Scheme (NEQAS) exercises and, because they are of no value in routine patient typing, their use is not recommended.10,11 Selection of high-avidity monoclonal anti-D reagents will allow detection of all but the weakest examples of weak D, negating the need to use more sensitive techniques to check the D status of apparent D negatives.


Some anti-D reagents have high levels of potentiator (e.g. polyethylene glycol) and should be used with caution; a diluent control is essential to demonstrate that the diluent does not promote agglutination of the test red cells as may happen if the patient’s cells are coated in vivo with immunoglobulin G (IgG); any positive reaction seen with the control, however weak, invalidates the test result. Anti-D reagents are provided in many different kits and those responsible for selection and purchase should make themselves aware of the content, specificity and potentiation of the chosen reagent.



Methods


There are several techniques available for routine ABO and D grouping including tube test, slide test, liquid-phase and solid-phase microplates and columns. Other techniques for blood grouping have been described but they are not in routine use. For example, molecular ABO typing is reserved for investigating anomalous ABO groups, in organ transplantation where red cells from the donor are not available, forensic practice and paternity testing.23 Care should be taken to use the appropriate reagent because not all reagents have been validated by the manufacturer for all techniques.



Tube and slide tests


Spin-tube tests may be used for urgent testing, where small numbers of tests are performed at once. Slide or tile techniques are widely used in under-resourced countries for ABO and D grouping. Spin-tube tests should be performed in 75 × 10 or 75 × 12 mm plastic tubes. Immediate spin tests may be used in an emergency, whereas routine tests are usually left for 15 min at room temperature (about 20°C) before centrifugation for 1 min at 150 g. Equal volumes (1 or 2 drops from either a commercial reagent dropper or a Pasteur pipette) of liquid reagents or plasma and 2% cell suspensions are used. The patient’s red cells (diluted in phosphate buffered saline, PBS) should be tested against monoclonal anti-A and anti-B grouping reagents. The patient’s plasma should be tested against A1 and B reagent red cells (reverse grouping). In addition, the plasma should be tested against either the patient’s own cells or group O cells (i.e. a negative control) to exclude reactions with A and B cells as a result of cold agglutinins other than anti-A or anti-B in the patient’s sample. Mix the suspensions by tapping the tubes and leave them undisturbed for 15 min. Agglutination should be read as described on p. 498. Any discrepancy between the results of the red cell grouping and the reverse grouping should be investigated further and any repeat tests should involve cells taken from the original sample rather than the prepared suspension. Reverse grouping is not carried out for infants younger than 4 months of age because the corresponding antibodies are normally absent or maternal in origin.





Liquid-phase microplate methods


Liquid-phase microplate technology provides a cheap and secure method for batch testing when semiautomation is utilized for dispensing and reading but it is no longer the grouping technique of choice in the UK (see Fig. 22.2). In 2009, a UK NEQAS survey showed that only 13% of responding laboratories were using microplates for grouping, down from 41% in a similar survey in 2002.10


ABO and D grouping may be performed in a single microplate if monoclonal reagents are used. A resuspension technique using untreated U-well rigid polystyrene microplates is recommended for grouping. If microplates are to be reused they should be cleaned in mild detergent, rinsed thoroughly in distilled water and left to dry face down; alternatively, an ultrasonic bath may be used. Scored or otherwise damaged plates should be discarded.


The plate is usually laid out as 12 × 8 (12 tests and 8 reagents) but may be used in the opposite orientation (8 × 12), depending on the number of reagents and controls required. Particularly if performing this technique manually, the anti-D reagents should be kept away from the anti-A and anti-B reagents because splashing between wells can occur when dispensing reagents and handling the plates during testing if insufficient care is taken. The following method is recommended:24












Column agglutination techniques


CAT techniques (see Chapter 21, p. 502) are now the commonest method for grouping in the UK (80% of laboratories who responded to a UK NEQAS survey in 2009, see Fig. 22.2), especially where automated systems are in place; these should always be performed in accordance with the manufacturer’s instructions. There are several different profiles to choose from and some cards/cassettes include monoclonal antibodies to other blood group antigens (e.g. K) in addition (see Chapter 21, p. 502 and Fig. 22.6). Forward and reverse grouping may be undertaken in separate cards/cassettes.





Controls


Positive and negative controls should be included with every test or batch of manual tests. In fully automated systems, the controls should be set up at least twice in a 24-h period. The timings should coincide with machine startup and changing of reagents, taking account of the length of time that reagents have been kept at room temperature on the machine. The control samples should be loaded in the same way as the test samples. The required controls are shown in Table 22.2. Where controls do not give the expected reactions, investigations should be undertaken to determine the validity of all tests undertaken subsequent to the most recent valid control results.


Table 22.2 Control cells for blood grouping



























Reagent Positive control Negative control
Anti-A A cells B cells
Anti-B B cells A cells
Anti-D D positive cells D negative cells
A1 cells Anti-A Anti-B
B cells Anti-B Anti-A


Causes of Discrepancies in ABO/D Grouping





T-activation/polyagglutination

Polyagglutination25 describes agglutination of red cells by all or most normal adult sera but not by the patient’s own serum. This is as the result of IgM antibodies reacting with an antigen on the red cells which is usually hidden but can be exposed by enzyme activity. The most common form is T-activation, which occurs when the bacterial enzyme neuraminidase cleaves N-acetyl neuraminic acid from the red cell membrane, exposing the T antigen.


This used to be a problem when grouping with polyclonal reagents, which contain anti-T, but tests using monoclonal reagents are not affected by this phenomenon.






False-Negative Reactions






Mixed-field appearance

This describes a dual population of agglutinated and non-agglutinated red cells which may be observed in both ABO and D grouping. It is important to recognize this as a mixed-field picture and not to confuse it with weak agglutination. The most likely cause of a mixed-field picture is the transfusion (either deliberate or accidental) of non-identical ABO or D red cells. Investigation will be required to determine the actual blood group of the patient, who may have been transfused in an emergency, at a different establishment or may have received an intrauterine transfusion. A mixed-field ABO group may be the first indication of a previous ABO-incompatible transfusion.


An ABO or D incompatible haemopoietic progenitor cell transplant will result in a mixed-field picture until total engraftment has occurred; the mixed-field picture may subsequently reappear when a graft is failing. Rarely, a dual population of cells is permanent and results from a weak subgroup of A (A3) or a blood group chimerism.


Interpretation of a dual population of red cells will depend on the technique used. In a tube, microscopic reading will reveal strong agglutinates in a background sea of free cells. In column agglutination techniques there will be a line of agglutinated cells at the top of the column, with the non-agglutinated cells travelling through to the bottom of the column. In liquid-phase techniques if the reaction grade is not a strong positive or an obvious negative then the reactions requires further investigation, which may include microscopic examination. In a solid-phase technique a mixed field is seen as a dual population of cells, with the agglutination being surrounded by free cells. Automated systems should be set up to detect mixed-field pictures and this should be used in conjunction with local policies.



D variant phenotypes

Weak D phenotypes are where the entire D antigen is present but there are fewer D antigen sites per cell and most weak D types group as D positive with the currently available high-avidity commercial monoclonal anti-D reagents. Where differing reactions are obtained with two reagents, the patient may be a partial D (i.e. one or more of the epitopes of the D antigen is missing). It was generally thought that patients who are weak D are unable to make anti-D and may be treated as D positive whereas some patients with partial D may be capable of making immune anti-D following sensitization with the missing epitope. This concept has been challenged by reports of weak D patients with anti-D.28


Detailed genotypic analysis combined with structural modelling of the D antigen in D variants have demonstrated change in the amino acid sequence in the extracellular domain in partial D, whereas weak D is associated with mutations resulting in change to the amino acid sequence in the intracellular or membrane-spanning domain.


A weak reaction with a single anti-D reagent should be investigated with a second anti-D reagent before assigning a result of D positive. It is essential to be able to distinguish between a weak reaction and a mixed-field reaction because the latter may be the result of a patient who is being transfused with blood of a different D type.


If in doubt, it is safer to call the patient D negative, at least until investigations have been undertaken by a reference centre. This will be of no clinical consequence because it is safe to transfuse D negative blood to a patient who is D positive, and a pregnant woman who is D positive and her unborn child would be unlikely to be harmed by the injection of prophylactic anti-D immunoglobulin.


Current advice about choice of D-typing reagents is that because DVI lacks the most epitopes, such individuals are likely to make anti-D when challenged by transfusion or pregnancy. For this reason, anti-D reagents for routine grouping of patients’ samples should not detect DVI.11 There is little evidence to suggest that a DVI donor would elicit an immune response in a recipient who is D negative; however, weak D positive and partial D donors, including DVI donors, should be classified as D positive.29 There are some important ethnic differences in the frequency of different Rh haplotypes, as shown in Table 22.3. Resolution of anomalous D grouping where a partial D is suspected now includes both serological testing and genotypic studies.30




Antibody screening


Antibody screening is usually undertaken at the same time as blood grouping and in advance of selecting blood for transfusion. Antibody screening may be more reliable and sensitive than crossmatching against donor cells because some antibodies react more strongly with red cells with homozygous expression (double dose) of the relevant antigen than with those with heterozygous expression (single dose) – most notably anti-Jka/Jkb but also anti-Fya, -Fyb, -S and -s. Screening cells can be selected to reflect this, whereas donor cells are usually of unknown zygosity. In addition, reagent red cells are easier to standardize than donor cells and there is potentially less opportunity for procedural error, particularly in automated systems.


Clinically significant antibodies are those that are capable of causing patient morbidity as a result of accelerated destruction of a significant proportion of transfused red cells. With few exceptions, clinically significant antibodies are those that are reactive in the indirect antiglobulin test at 37°C; however, it is not possible to predict serologically which of these antibodies will definitely be of clinical significance, so the term ‘of potential clinical significance’ is often used.



Red Cell Reagents


The patient’s plasma should be tested against at least two individual screening cells, used individually, not pooled. The screening cells should be group O and encompass the common antigens of the indigenous population.


In the UK, the following antigens should be expressed as a minimum: C, c, D, E, e, K, k, Fya, Fyb, Jka, Jkb, S, s, M, N and Lea; one cell should be R2R2 and another R1R1 or R1wR1. The following phenotypes should also be represented in the screening set: Jk(a+b−), Jk(a−b+), S+s−, S−s+, Fy(a+b−) and Fy(a−b+) (Table 22.4). These recommendations for homozygosity are based on UK data regarding the incidence of delayed haemolytic transfusion reactions, the need for high sensitivity in the detection of Kidd antibodies and the poorer performance of column agglutination techniques in the detection of some examples of Kidd antibodies using heterozygous cells.11,29 The requirement for the expression of Cw and Kpa antigens on screening cells has been the cause of much debate but in the UK and the USA detection of anti-Cw or anti-Kpa is not a requirement even in the absence of an antiglobulin crossmatch.11,29 This is because these are low-frequency antigens and the antibodies rarely cause delayed haemolytic transfusion reactions or severe haemolytic disease of the newborn.


Table 22.4 Expression of red cell antigens on screening cells11













Only gold members can continue reading. Log In or Register to continue

Stay updated, free articles. Join our Telegram channel

Jun 12, 2016 | Posted by in HEMATOLOGY | Comments Off on Laboratory aspects of blood transfusion

Full access? Get Clinical Tree

Get Clinical Tree app for offline access
Blood group system Antigen Homozygous cells recommended
Rh C Yes (R1R1 or R1wR1)